2-18 In-Band/Out-of-Channel Measurements
Spectrum Analyzer
2-24
PN: 10580-00349 Rev. H
Spectrum Analyzer MG
2-18 In-Band/Out-of-Channel Measurements
The in-band/out-of-channel measurements are used to determine the distortion and
interference within the system band, but outside of the transmitting channel. These
measurements include in-band spurious emissions and adjacent channel power ratio (also
called spectral regrowth). There are stringent regulatory controls on the amount of
interference that a transmitter can spill to neighboring channels. To determine compliance
with the allowable level of spurious emissions, two parameters need to be specified:
•
Measurement channel bandwidth
•
Allowable level of spurious emissions
The measurements described in this section are distinct from the built-in emission mask
(in-band) measurements that are performed via the
Masks and C/I
Required Equipment
•
30 dB, 50 watt, Bi-Directional, DC–18 GHz, N(m)–N(f), Attenuator,
Anritsu 42N50A-30
•
Test Port extension cable, Anritsu 15NNF50-1.5C
Procedure
1.
Using the test port extension cable and 30 dB, 50 watt, (Bi-directional) attenuator,
connect the
RF In
port to the appropriate transmit test port.
2.
Press the
Freq
main menu key, press the
Center Freq
submenu key, and enter the
center frequency.
3.
Press the
Span
main menu key. Set the span wide enough to include the primary
channel bandwidth and upper and lower channel bandwidths.
4.
Press the
Amplitude
main menu key and then press the
Reference Level
submenu key
to set the reference level to
–20 dBm
.
5.
Press the
RL Offset
submenu key, enter
30
then select dB
External Loss
to compensate
for the loss in the attenuator.
6.
Press the
Auto Atten
submenu key and set the attenuation to
On
.
7.
Press the
BW
main menu key and use the
RBW
and
VBW
submenu keys to set the
resolution bandwidth to
10 kHz
and the video bandwidth to
300 Hz
.
8.
Press the
Marker
main menu key and press the
Marker 123456
submenu key to select
Marker
1
. The underlined number indicates the active marker.
9.
Press the
On/Off
submenu key and use the arrow keys, the keypad or the rotary knob to
move the marker over one of the spurs.
10.
Compare the value of the marker to the specified allowable level of
in-band/out-of-channel spurious emissions for the corresponding channel transmit
frequency.
11.
Repeat steps 9 and 10 for the remaining spurs. Use either Marker 1 again, or choose
another marker.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...