2-3
Measurement Example
Chapter 2 — E-Series Tracking Generator
2-4
PN: 10580-00339 Rev. C
Tracking Generator MG
Enhancing Accuracy
For detailed power level measurements, the output flatness of the tracking generator can be
calibrated in the field.
1.
After the frequency range and other settings are determined, disconnect the DUT and
make a through connection between the
GENERATOR/RF OUT
connector and the
ANALYZER/RF IN
connector. For the most accurate results, use the same cables and
adapters for this connection as were used to connect the DUT.
2.
From the
Generator > Setting
submenu, press
Enhanced Generator Power Accuracy
.
3.
Select the number of external
Matching Pad (3 dB)
pads to use for calibration and
measurements. Press
Enhanced Accuracy Calibration
. External attenuators will need to
be inserted for the
3 dB
and
6 dB
pad options.
4.
Follow the onscreen instructions shown in
During the Enhanced Accuracy Calibration process, the instrument calibrates across
the frequency range, reducing any output ripple. After the successful completion, the
(
+Accy
) label will be displayed after the Generator output power (
).
5.
is a detail of the top of the filter trace shown in
calibration.
shows the same trace section after enhanced calibration using
two 3 dB pads.
The 7.5 dB loss from the 0 dBm generator power is from the two 3 dB pads, the filter,
and cables connected between the Generator port and Spectrum Analyzer receiver.
Note
Matching 3 dB external attenuators are not required. Attenuators do provide a
more accurate 50 ohm match to the generator output and analyzer input. This may
help in cases where transmission ripple needs to be minimized.
For the most accurate results, use the same 3 dB pad setup (0, 1, or 2 pads) in
measurements as were used for calibration.
Caution
The Enhanced Accuracy calibrations are eliminated at Preset, Power-on, or when
the generator output level is changed. The calibrations are not saved nor recalled
with Save/Recall.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...