Connecting R&S TSMW to Nemo Invex with an Ethernet
Cable
Rohde&Schwarz TSMW scanner is connected to the Nemo Invex chassis (SI module) with an
Ethernet cable.
After you have connected the TSMW scanner to Invex, you will need to change the IP address
of Nemo Invex to match the IP address of the scanner. This requires a USB memory key.
1. Create a folder named
Nemo Invex
on the USB memory key.
2. Create a new file named
SetIPAddress.txt
.
3. Open the txt file and type the new IP address in the following format
Base_IP,Subnet mask,Gateway
4. For example, the default Nemo Invex IP address must be changed from 192.168.3.210 to
192.168.0.3. Type
192.168.0.3,255.555.254.0,192.168.1.1
. Save and close the txt file.
5. Connect the USB memory key to the USB-1 port on the UIC 1 module of your Nemo Invex
chassis.
6. Wait until red LEDs are blinking and the shut down the chassis.
7. Remove the USB memory key.
8. The new IP address is assigned before the Nemo Measurement server starts up.
9. For other UICs IP address is incremented by one automatically.