Maintenance
7.1. Introduction
This chapter contains maintenance instructions for the Accent XL system.
Routine maintenance may be performed by clinic staff unless otherwise specified.
Any maintenance procedure not mentioned in this chapter must be performed only
by Alma Lasers authorized technical personnel.
The system is designed to operate reliably without any need for operator
maintenance. However, the outer surfaces of the system should be kept clean for
hygienic reasons and the applicator tips should be cleaned between sessions to
enable effective treatment.
Warning
Maintenance by the operator should be performed only when the system is shut
down and disconnected from the mains power source. Performing maintenance
procedures with the system turned on may be hazardous to the operator and/or
destructive to the system.
7.2. Periodic
Service
The system should be periodically inspected and maintained to keep it in peak
operating condition.
The following routine service operations should be performed by Alma Lasers
authorized technical personnel every 6 months:
•
General system check
•
Interior inspection, including cleaning accumulated layers of dust
•
Check the radiofrequency output energy
•
Cooling system:
Check the water level in the cooling system; refill if necessary
Replace the deionizer filter
Clean the radiator
Check the water quick-connector O-rings on the handpieces
Warning
•
The Accent XL system generates hazardous voltages within the main
console.
•
The interior of the system may be serviced only by Alma Lasers authorized
technical personnel.
7-2 Accent XL
™
Operator's Manual