Troubleshooting
Table 8-1: Error Message Troubleshooting Guide (continued)
Error Message/Alarm
Probable Cause
Corrective Action
Handpiece Changed
The handpiece was replaced
while the system was turned on
1. Turn system off, wait 1
minute and restart; resume
normal operation
2. If problem persists, call Alma
Lasers Service
Error Wiring Data
Data cannot be saved
1. Turn system off, wait 1
minute and restart; resume
normal operation
2. If problem persists, call Alma
Lasers Service
Flow SW Error
Insufficient coolant flow in the
system
1. Add deionized water to the
cooling system
2. If problem persists, call Alma
Lasers Service
Table 8-2: System Malfunctions Troubleshooting Guide
Symptom
Probable Cause
Corrective Action
Power disconnected
•
Check power cable
•
Plug cable into wall
Red emergency shutoff knob is
engaged
Release emergency shutoff
knob by turning it clockwise
System does not start
Burnt fuse
Replace the fuses
Make sure the power cable in
the rear of the monitor is
connected
Monitor does not respond when
the system is on
The monitor is not receiving
power
8-4 Accent XL
™
Operator's Manual