Clinical Guide
Table A-3: Perioral Treatment Settings
Accent XL Handpiece
Treatment
Pass #
Pass Duration
[Seconds]
Target
Tissue Temp.
(°C)
Unipolar
Bipolar
1
30
39 – 40
90 – 100 W
Pass # 1 N/A
2
30
39 – 40
90 – 100 W
10% reduction as needed
from Pass # 1 to maintain
target temperature and
patient comfort
Pass # 2 N/A
50 – 60 W
10% reduction as needed
from Pass # 2 to maintain
target temperature and
patient comfort
3
30 (UP)
20 (BP)
41 – 43
85 – 100 W
10% reduction as needed
from Pass # 2 to maintain
target temperature and
patient comfort
4
30 (UP)
20 (BP)
41 – 43
85 – 100 W
10% reduction as needed
from Pass # 3 to maintain
target temperature and
patient comfort
45 – 55 W
10% reduction as needed
from Pass # 3 to maintain
target temperature and
patient comfort
5
30 (UP)
20 (BP)
41 – 43
85 – 100 W
10% reduction as needed
from Pass # 4 to maintain
target temperature and
patient comfort
45 – 50 W
10% reduction as needed
from Pass # 4 to maintain
target temperature and
patient comfort
6
30 (UP)
20 (BP)
41 – 43
85 – 100 W
10% reduction as needed
from Pass #5 to maintain
target temperature and
patient comfort
45 – 50 W
10% reduction as needed
from Pass #5 to maintain
target temperature and
patient comfort
Accent XL
™
Operator's Manual A-13