- Dokumentation
Maintenance
System Manual
Cooling water system
CRIUS II
6
6.14
Cooling water system
6.14.1 Checking the water pressure
1
Open the reactor cabinet to access the cooling water system. For iden-
tifying the components please refer to the delivered cooling water dia-
gram.
Fig. 6-38
Cooling water system
A Flow meter with flow limiter
B shut-off valve
C Water filter
D Pneumatic valve
E Manual valve
Checking the inlet
pressure
2
Check the pressure display at the filter units
•
You find the required pressure values in the delivered installation
plan
3
If necessary, adjust the pressure by means of the handwheel
•
If the required pressure cannot be reached this indicates a contam-
inated filter. See Chapter 6.14.2 Cleaning the water filter,
Checking the outlet
pressure
4
Check the pressure display at both outlets
•
You find the required pressure values in the delivered installation
plan