Maintenance
System Manual
Pneumatic system
CRIUS II
6
6.12.3 Checking the pressure sensor
The pressure sensor causes an alarm if the working pressure decreases
below the set limit.
1
Check the pressure limit value at the pressure sensor
•
The pressure limit must be set to 4 bar
2
If necessary, adjust the pressure limit.
See OEM manual.
Fig. 6-32
Pressure sensor
A Pressure display
B Operating elements
6.12.4 Checking the EPV valves
1
Activate the EPV valve to be checked in AIXACT
•
The valves are labeled!
2
Check the activated EPV valve at the corresponding valve cluster
•
The LED at the activated valve must emit light
Fig. 6-33
Valve cluster
A EPV valves
B Pneumatic hoses