Maintenance
CRIUS II
Cooling water system
Fig. 6-41
Adjusting the flow
A Manual valve
Checking the flow
limits
6
Check the upper and lower limit flow settings in the Data Sampler
•
If a flow value is out of range an alarm occurs
7
If necessary, correct the limit values in the Data Sampler
6.14.4 Unintended release of liquid
The entire system is equipped with liquid sensors which are connected to the
SCS.
Fig. 6-42
Liquid sensor
If a sensor detects liquid (e.g. cooling agent from a thermostat heat bath), the
SCS automatically triggers the required safety measures.
1
Remove the liquid with a vacuum cleaner which is suitable for liquids
and eliminate the leak