118
Chapter 7
Characteristics
General Characteristics
C
h
ar
ac
ter
is
ti
cs
General Characteristics
Power
Environmental
Samples of this product have been type tested in accordance with the Agilent
Environmental Test Manual and verified to be robust against the environmental
stresses of Storage, Transportation, and End-use; those stresses include but are not
limited to temperature, humidity, shock vibration, altitude, and power line
conditions. Test methods are aligned with IEC 60068-2 and levels are similar to
MIL-PRF-28800F Class 3.
Safety
Designed for compliance to IEC 61010-1:2001
EMC
Meets the conducted and radiated emissions and immunity requirements of
IEC 61326:2002 when tested with EMC shielded filler panels (Agilent P/N
N6030-80007, kit of 6) separating the controller and the N6030A module,
and in all open slots. The RFI gaskets must be oriented to the right.
Supply
Typical Operation (Watts)
+3.3 VDC
11.2
+5 VDC
22
+12 VDC
5
-12 VDC
5
Total Power
43.2
Operating temperature:
0 to +55 degrees C
Storage temperature:
-20 to +70 degrees C
Relative Humidity:
Type tested:
10 to 90% at 40 degrees C (non-condensing)
Altitude:
0 to 2000m (6500 ft) above sea level