Chapter 1
13
Introducing the N6030A Series AWGs
The N6030A Series Module
Intro
d
u
cing
th
e N60
30A Series
AW
G
s
CH 1/CH 2 OUT
The CH 1 OUT and CH 2 OUT positive (+) connectors are used for
single-ended operation. Use both the positive (+) and negative (-) connectors
for differential operation. Refer to
“Signal Conditioning” on page 75
for
more information.
DATA PORT
This port is currently inactive and reserved for future applications.
AUX PORT
The AUX port enables Dynamic Sequencing, Option 300.
EXT CLK IN
Use this 50 ohm SMA external clock in connector to input an external
sample clock. It will accept clock rates in the range of 100 MS/s through
1.25 GS/s. Refer to
“External Clock” on page 61
for more information.
NOTE
An error message will appear if the clock rate does not match the hardware
setting, or an external clock is not present.
INT CLK OUT
Use this 50 ohm SMA internal clock out connector to route the internal 1.25
GS/s clock to other test instruments or devices.
10MHz REF IN
Use this 50 ohm SMA 10 MHz reference in connector to input an external
10 MHz reference.
SYNC CLK IN/SYNC CLK OUT
These connectors support synchronization of multiple modules. Refer to
“Multiple Module Synchronization” on page 78
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com