5-20
Service Guide E8364-90038
Theory of Operation
PNA Series Microwave Network Analyzers
Signal Separation Group Operation
E8362C, E8363C, E8364C
swapping the front panel jumpers for one port, signal flow through the corresponding coupler is reversed,
increasing the test signal sensitivity by 15 dB.
In the forward direction, as shown in
Figure 5-5
, the signal flow through the test port coupler (A26) is
reversed by arranging the front panel jumpers such that RCVR B IN connects to CPLR THRU and CPLR ARM
connects to SOURCE OUT.
In the reverse direction, not shown, the signal flow through the test port coupler (A25) is reversed by
arranging the front panel jumpers such that RCVR A IN connects to CPLR THRU and CPLR ARM connects to
SOURCE OUT.
For more information on high dynamic range measurements and configurations, search for “Option 014” in
the embedded help index in the analyzer.
Figure 5-5
High Dynamic Range Configuration in the Forward Direction