Tests and Adjustments
E8362C, E8363C, E836C
Adjustments
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network analyzer’s 10
MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the Frequency Adjustment at 10 MHz
Procedure
NOTE
The specification of frequency accuracy is
1 ppm. However, this test typically adjusts to
0.01 ppm.
1. Connect the equipment as shown in
Figure 3-17
. Connect a GPIB cable between the network analyzer
and the frequency counter.
Figure 3-17
Setup for Adjustment of the 10 MHz Frequency Reference
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
10 MHz Freq Adjust
.
3. Ensure the GPIB settings are correct.
4. Click
Begin Adj
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model or
Part Number
Cable, BNC, 50
, 24 inch
All
8120-1839
Any
Frequency counter
All
53151A
Opt 001
Any
a
a. Must be capable of measuring a signal at 10 MHz.
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...