Service Guide E8364-90038
4- 35
PNA Series Microwave Network Analyzers
Troubleshooting
E8362C, E8363C, E8364C
Measurement System Troubleshooting
6. If the signals are <-30 dBm or not present, replace the A17 LOMA 10. Refer to
“Removing and Replacing
the A17 LOMA 10” on page 7-34
.
7. If the signals are present, reconnect cable W18, and then continue with
“Checking the A18 MA 20
Output” on page 4-37
.