4-18
Service Guide E8364-90038
Troubleshooting
PNA Series Microwave Network Analyzers
Rear Panel Troubleshooting
E8362C, E8363C, E8364C
Checking the SERIAL (RS-232), PARALLEL (1284-C), or VGA Port
To verify the proper operation of the SERIAL, PARALLEL, or VGA port:
•
Connect a known good serial, parallel, or VGA peripheral device.
•
Wait 15 seconds for the analyzer to verify the device connection, and then check the operation of the
peripheral device.
•
If the peripheral device performs correctly, the port is functioning properly.
•
If the peripheral device does not function properly, the A15 CPU board is faulty. Refer to
“Removing and
Replacing the A15 CPU Board” on page 7-28
.
Checking the GPIB Port
The network analyzer uses a National Instruments 488.2 GPIB controller and associated driver software.
This software includes a test utility which scans the GPIB bus and returns the status of all the connected
peripherals.
To run the test utility software and check the GPIB status:
1. Connect a known good peripheral to the analyzer using a known good GPIB cable.
2. On the
System
menu, point to
Configure,
and then click
SCPI/GPIB
.
A
SCPI/GPIB
dialog box is displayed.
3. In the
GPIB
block, click
System Controller
to establish the analyzer as a controller. Wait for the analyzer to
configure, and then click
OK
.
4. On the
System
menu, click
Windows Taskbar
to
open the
Start Menu
window.
5. On the
Start Menu
window, point to
Programs, National Instruments NI-488.2,
and then click
Explore GPIB
to
open
the
Measurement & Automation
window.
6. On the left side of the
Measurement & Automation
window under folders:
a. Click the plus sign to expand the
Measurement & Automation
folder.
b. Click the plus sign to expand the
Devices and Interfaces
folder.
c. Right click
GPIB0 (AT-GPIB/TNT)
to open a submenu.
7. On the submenu, click
Scan for Instruments
to run the test.
8. The state of all the peripherals found on the bus is returned.
9. If problems are detected, check the connections of all GPIB cables, and check all the GPIB addresses of
the instruments on the bus.
NOTE
Address Information
•
Each device must have its own unique address.
•
The network analyzer’s default GPIB address in the controller mode is 21.
•
The address set on each device must match the one recognized by the analyzer (and
displayed).
Refer to the manual of the peripheral to read or change its address.
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...