8-14
Service Guide E8364-90038
General Purpose Maintenance Procedures
PNA Series Microwave Network Analyzers
Error Terms
E8362C, E8363C, E8364C
Isolation (E
XF
and E
XR
)
Isolation, or crosstalk, is the uncorrected forward and reverse isolation error terms that represent leakage
between the test ports and the signal paths. The isolation error terms are characterized by measuring
transmission (S
21
, S
12
) with loads attached to both ports during the measurement calibration.isolation
errors affect transmission measurements primarily where the measured signal level is very low.
The measurements most affected by isolation error terms are DUTs with large insertion loss. Since these
terms are low in magnitude, they are usually noisy (not very repeatable).
Typical Cause of Failure
A
loose cable connection or leakage between components
in the test set are the
most likely cause of isolation problems.
After verifying the cable and its connections, suspect the analyzer
switch splitter, step attenuator, coupler,
or receivers, and associated cabling
.
To troubleshoot, refer to
“Checking the Receiver Group” on page 4-42
or to
“Checking the Signal Separation
Group” on page 4-38
.
Summary of Contents for E8362C
Page 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Page 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Page 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Page 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Page 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Page 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Page 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...