The instrument allows you to combine an external measurement device
such as a Digital Multimeter (DMM) with multiplexer channels to create
a scan. During a scan, the instrument closes the configured multiplexer
channels one at a time to allow a measurement to be made on that
channel.
Before initiating a scan, a scan list must be set up. You can also specify
an arm source, a trigger source, and the number of sweeps (a sweep is
one pass through the scan list) to control the scan process. All these can
be done from the S.List menu. The procedure in this chapter describes a
simple scan from the front panel. For more information about scanning
and using the parameters to control a scan, see page 85.
To Create a Scan List
1. Press the S.List key to enter the scan list menu. The CONFIG
annunciator lights up and the first level of the menu is displayed.
2. Press Enter to begin the channel selection.
3. Turn the knob to select the first channel for the list (e.g., 103).
4. Press Enter to add the channel into the scan list. The starburst
character lights in the display to indicate the channel is now a part of
the scan list.
ADD TO SCAN
101
SELECT
101
SELECT
103
SELECT *
103
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...