166
6
Error Messages
•
Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. Errors are cleared as you
read them. When you have read all errors from the queue, the
ERROR annunciator turns off and the errors are cleared. The
instrument beeps once each time an error is generated.
•
If more than 10 errors have occurred, the last error stored in the
queue (the most recent error) is replaced with
-350, “Queue overflow”
.
No additional errors are stored until you remove errors from the
queue. If no errors have occurred when you read the error queue, the
instrument responds with +0, “
No error
”.
•
The error queue is cleared by the
*CLS
(clear status) command or
when power is cycled. The errors are also cleared when you read the
queue. The error queue
is not
cleared by an instrument reset (
*RST
command) or a card/module reset (
SYSTem:CPON
command).
•
Front-Panel Operation:
If the
ERROR
annunciator is on, press View, select ERROR, and
press Enter. Use the knob to scroll through the error numbers. Press
the arrow key (right) to view the text of the error message. All errors
are cleared when you exit the menu.
•
Remote Interface Operation:
SYSTem:ERRor?
Read and clear one error from the error queue.
Errors have the following format (the error string may contain up to
80 characters):
-113, “Undefined header”
01:ERR -113
First error in queue
Error code
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...