88
Chapter 3, Programming the PCS Interface using GPIB Control
Set Up and Measure
NOTE:
If the Immediate trigger is used, up to 48000 samples can be collected by the combination
of the LENGTH and AVERAGE values.
1. The total number of points used in the power measurement determines the accuracy achieved.
The measurement accuracy is:
Where:
N = total number of points used, or LENGTH*AVERAGE from the table above
2. The PCS Interface’s power detector sample rate is 80 kHz. For the US TDMA symbol rate of
24.3 kHz, this implies 3.29 samples/symbol. The values for IGNORE and LENGTH provide
power measurements compliant with the IS-137 specification of at least 140 symbols between
symbols 6 and 162 in the burst.
Table 9
Example power measurement set ups for typical applications
TX:INP:POW:PDET:SAMP
TX:INP:POW:TRIG
:IGN
(IGNORE)
:LENG
(LENGTH)
:AVER
(AVERAGE
)
TRIGGERING
TYPE
A
pplica
tio
n
CDMA
Base Station
—
4800
4
IMM (Immediate)
CDMA
Subscriber Unit
—
4800
4
IMM (Immediate)
TDMA
Base Station
—
See footnote
1
1
IMM (Immediate)
TDMA
Subscriber Unit
2
See
footnote
3
461
EXT (External)
TDMA
Subscriber Unit
(Signals >
−
5 dBm)
17
461
AMPTD (Amplitude)
± 5%
180
N
----------
µ
W
[
]
This applies @ 23 ± 10 degrees C
±
± 10%
180
N
----------
µ
W
[
]
This applies @ 0 to 55 degrees C
±
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...