Chapter 2, Installation
When Used with Agilent 8924C and Controlled via Serial Port for CDMA Subscriber Unit
Tests
Cable Connections
Follow these steps to prepare the test equipment for use:
1
Set the GPIB Address Selector of the rear panel of the PCS Interface as follows.
GPIB/Ser switch
Ser
GPIB address
Any
2
Make the front and rear panel connections shown in figure 5.
3
Turn on the PCS Interface.
NOTE:
The PCS Interface must be powered on and connected before the Agilent 8924C is powered
on. If the PCS Interface is not powered on first, the Agilent 8924C will not detect it.
4
Set up the Test Set.
a
Turn on the Test Set.
b
Press and release the SHIFT key then the TESTS key to access the CONFIGURE
screen.
c
Position the cursor in the “PCS Intrfc Control” field, and press the knob to select
“On.”
d
Turn the Test Set’s power off and then back on.
5
Perform the RF Generator Level Calibration.
a
Press and release the SHIFT key then the TESTS key to access the CONFIGURE
screen.
b
Position the cursor in the “RF Gen Level” field, and press the knob to start the
calibration.