136
Chapter 5, Specifications
General Characteristics
Others
Safety:
Complies with IEC 61010-1:1990 +A1:1992 +A2:1995 / EN 61010-1:1993
+A2:1995
Certified by CAN / CSA C22.2 No. 1010.1-92
EMC:
Complies with IEC 61326-1:1997 +A1:1998 / EN 61326-1:1997 +A1:1998
Standard
Limit
CISPR 11:1990 / EN 55011:1991 /
AS/NZS 2064.1/2
Group 1, Class A
IEC 61000-4-2:1995 / EN 61000-4-2:1995
4 kV CD, 8 kV AD
IEC 61000-4-3:1995 / EN 61000-4-3:1996
3 V/m 80% AM 27 - 1000 MHz
IEC 61000-4-4:1995 / EN 61000-4-4:1995
0.5 kV signal lines, 1 kV power lines
IEC 61000-4-5:1995 / EN 61000-4-5:1995
0.5 kV line-line, 1 kV line-ground
IEC 61000-4-6:1996 / EN 61000-4-6:1996
3 V 80% AM 0.15 - 80 MHz
IEC 61000-4-11:1994 / EN 61000-4-11:1994
1 cycle, 100%
NOTE:
When tested at 3 V/m according to IEC 61000-4-3:1995, the output level accuracy will be
within specifications over the full immunity test frequency range of 27 to 1000 MHz except
when the analyzer frequency is identical to the transmitted interference signal test
frequency.
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...