52
Chapter 2, Installation
When Used with Agilent 8921A for CDMA Base Stations Tests
Cable Connections
Follow these steps to prepare the test equipment for use:
NOTE:
You should have made all connections between the Test Set and Cellular Adapter at this
point. Refer to the User’s Guide for your particular Cellular Adapter for instructions on
connecting it to the Test Set. Verify these connections before continuing with these steps.
1
Set the GPIB Address Selector on the rear panel of the PCS Interface as follows.
GPIB/Ser switch
GPIB
GPIB address
19 (for the manual control software and connectivity test)
2
Make the front and rear panel connections shown in figure 15.
Consult the User's Guide for your Cellular Adapter for further details about using
these timebase signals in your measurements.
3
Turn on the PCS Interface.
NOTE:
The PCS Interface must be powered on and connected before the Agilent 8920B is powered
on. If the PCS Interface is not powered on first, the Agilent 8921A will not detect it.
4
Turn on the Test Set.