26
Chapter 1, General Information
Rear Panel
Rear Panel
Figure 4
Rear Panel
1. GPIB Address
Selector
The GPIB address selector specifies the GPIB interface address. The address
selector consists of the switches A1 to A5. The switch A1 is the least significant
bit (1) and A5 the most significant bit (16). For example, you may set the switches
A1, A2, and A5 ON (1) to specify 19 as your GPIB address.
The
GPIB/Ser
switch sets whether the GPIB port or SERIAL PORT is used to
control the PCS Interface.
2. REF IN
Reference frequency input port. The REF IN port connects to the 10 MHz
reference source. (Connector type: BNC (F))
3. REF OUT
Reference frequency output port. The REF OUT port may be used as a 10 MHz
source for other equipment. (Connector type: BNC (F))
4. EXT TRIG IN
External trigger input. EXT TRIG IN is used to trigger power measurements of
TDMA bursted signals. It connects to the FRAME CLK OUT port of the Cellular
Adapter when the UUT is the TDMA type. (Connector type: SMC (M))
5. DET OUT
Detector output. DET OUT can be used for power measurements of CDMA
subscriber units. (Connector type: SMC (M))
1 2 3 4 5
6
7
8
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...