101
Chapter 4, GPIB Commands
COMPensation Subsystem
COMPensation:PDET:DCOFfset:REQuest:RESolution
ëû
<
numeric
>[DEG]
This command specifies the threshold value of the temperature change used to
judge whether the update of the dc offset values of the power detector is required.
<numeric>:
Greater than or equal to 0 in 0.1°C steps.
If 0 is set, the temperature compensation is always required.
•
Response to the Query
{numeric} <new line><^END>
COMPensation:PDET:TEMPerature:REQuest:RESolution
ëû
<
numeric
>[DEG]
This command specifies the threshold value of the temperature change used to
judge whether the update of the voltage-power characteristics of the power
detector is required.
<numeric>:
Greater than or equal to 0 in 0.1°C steps.
If 0 is set, the temperature compensation is always required.
•
Response to the Query
{numeric} <new line><^END>
COMPensation[:RFG1]:TEMPerature:EXECute?
This command performs the generator level temperature compensation, and it
returns the signal path to the previous setting. (Query only)
NOTE:
Before executing this command, execute the
COMP:TEMP:FREQ?
ëû
{0|1|2|3|4|5}
and
the
COMP:TEMP:LEV?
command and input the specified frequency and signal level to
the FROM DUPLEX OUT port.
•
Response to the Query
{0|1} <new line><^END>
0:
Executed the temperature compensation but did not re-write the data,
because the measurement data values exceeds the internal limits. The
possible reason is that the cable connections are not correct or that the
level of the input signal to the FROM DUPLEX OUT port is not
correct.
1:
Executed the temperature compensation and re-wrote the data.
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...