156
Appendix C, Error Messages
Error Message
−
120
Numeric data error
The cause of the error that was generated was numeric data (which may include numeric
data that does not include a decimal point). This message is displayed if an error other
than those corresponding to error codes
−
121 to
−
129 is generated.
−
121
Invalid character in number
Data other than numeric data was contained within numeric data.
−
123
Exponent too large
An exponent was too large. The largest allowable exponent is 32000. (See IEEE 488.2,
7.7.2.4.1.)
−
124
Too many digits
Numeric data was too long. (The maximum allowable length is 255 digits.)
−
128
Numeric data not allowed
Numeric data may not be used.
−
130
Suffix error
A unit suffix error.
This message is displayed if a unit suffix error other than those corresponding to error
codes
−
131 to
−
139 is generated.
−
131
Invalid suffix
The unit suffix could not be recognized or was invalid.
−
134
Suffix too long
The unit suffix was too long.
This error is displayed if the unit suffix is 12 characters or more. (See IEEE 488.2,
7.7.3.4.)