![Acuster BEAT HF User Manual Download Page 61](http://html1.mh-extra.com/html/acuster/beat-hf/beat-hf_user-manual_2845321061.webp)
USER MANUAL
BEAT HF
Edition: January 2016
Revision: No. 4
MAINTENANCE -
61
The download can be stopped at any given time by pressing <STOP>: the entries that
have been downloaded up to that stage will have been exported to the file created on the
Pendrive.
Once the data has been exported (for further details related to the exportation, refer to
OPTION PENDRIVE-USB/A in section 3.5.3 of this Manual), the software version update
is begun automatically.
dd/mm/yy
hh:mm
Updating Firmware
Don’t turn off
The backlight of the screen goes off. The update
process lasts about a minute.
Once the download is complete, the unit will start up automatically (when the software
is updated with a new version the default language -English- is restored and existing
fusion records, fusion counter, operator and job data are erased).
WITH A COMPUTER VIA THE SERIAL PORT (not for model MV):
To carry out a download via the serial port of a computer, it must have the necessary
software installed. If it is the first time, install the Beat-HF Firmware Updater software
on the PC (only available in English).
Software downloading procedure:
1.
Connect the ODS transmission cable (model 2008) or Null Modem cable (model
2015) between the unit and the PC serial port.
2.
Execute the software on the PC. If it is the first time, choose the correct COM port
where the connection is made.
3.
Open the tab file (the file of the new beat-hf.bin software version to be download
should be copied), find and click the file to be loaded.
4.
Plug in the unit to mains and turn the master switch on while the
©
key is kept
pressed. The following message is displayed:
dd/mm/yy
hh:mm
Update Firmware?
<Yes> <No>
Confirm
Confirmation that the software version has been
updated is requested. The cursor remains situated on
<Yes> by default. To exit, press the
¨
arrow to move
it to <No> and press <START>. The unit will show the
first job screen.
With the cursor on <Yes> and pressing <START> the display will show the following
screen:
dd/mm/yy
hh:mm
Download fusions?
<Yes> <No>
Confirm
To update the software version via the serial port,
press the
¨
arrow to move it to <No> and press
<START>. The screen will switch to Update Firmware
directly.