- MODE OF USE
Revision: No. 4
28
nnnn
hh:mm
Power supply error
vvv V ff Hz tt
/
C
Fusion impossible
Appears when the input voltage or frequency is beyond
that tolerance:
Maximum limit: 270 Vac / 70 Hz
Minimum limit: 180 Vac / 40 Hz
Check the power source to correct the fault. It is not necessary to turn the unit on and off
using the main switch to refresh the initial value since if the minimum limit is attained,
the unit restarts at 200 Vac and 45 Hz (as appropriate) and if the maximum limit is
reduced, the unit restarts at 250 Vac and 65 Hz (as appropriate).
In case the voltage and frequency displayed on the screen were 0V and 0Hz, there is an
internal communication error. Contact to the Grupo Acuster’s After-Sales Service.
C
Room temperature beyond the limits:
If a room temperature beyond the established limits is detected (in the standard setup
they are -15ºC and +50ºC) or if the temperature probe has broken down, the following
message will appear when the unit is turned on:
dd/mm/yy
hh:mm
Wrong temperature
XXX
/
C
Confirm
When turned on, the unit will emit two acoustic signals
and a Wrong temperature message will appear on
the screen. xxx corresponds to the room temperature
read or if it is 255 it means the probe has broken down
or is not connected.
Press <START> to accept. The screen shown will correspond to the manual input of data
(using the keyboard).