Regular monitoring of the I
MPLANT
C
ENTER
unit is
necessary in order to detect any problem.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
11. 5 MICRO-MOTOR CORD AND SCALER CORD
MAINTENANCE
Important:
Do not use an abrasive product to clean the
cords.
The cords must not be immersed.
The cords must not be disinfected using
products containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected
and sterilized after each procedure, using
alcohol, disinfectant agents or disinfectant wipes
for dental practices, such as S
EPTOL
™ wipes.
b) Sterilization:
The cords can be sterilized in an autoclave under
the following conditions:
- Autoclave: Class B.
- Sterilization temperature: 134 °C.
- Time at sterilization temperature: 18 minutes.
11. 6 I-SURGE MICRO-MOTOR MAINTENANCE
Important:
Do not use an abrasive product to clean the I-
S
URGE
micro-motor.
The I-S
URGE
micro-motor must not be immersed.
The I-S
URGE
micro-motor must not be
disinfected using products containing acetone,
chlorine or bleach.
The I-S
URGE
micro-motor must not be cleaned in
an ultrasonic bath.
The I-S
URGE
micro-motor must not be
disassembled.
The I-S
URGE
micro-motor is a high-technology
device. To maintain optimal performance, the
following recommendations must be complied
with.
a) Cleaning and disinfection:
Use cleaning and disinfection products
compatible with the I-S
URGE
micro-motor (such as
S
PRAYNET
®
).
When using a spray, spray the outside and inside
of the I-S
URGE
micro-motor moderately to expel
residues.
Clean and disinfect the surfaces using a clean,
lint-free cloth impregnated with the product.
b) Lubrication:
Lubricate before each sterilization or at least
twice a day.
Use lubricants compatible with the I-S
URGE
micro-
motor (such as L
UBRIFLUID
®
).
Place a clean cloth over the I-S
URGE
micro-motor
ventilation vents to absorb the surplus lubricant.
Inject the lubricant through the nose of the
micro-motor using an oilcan of lubricant (or a
spray) for one-half to one second.
c) Sterilization:
The I-S
URGE
micro-motor can be sterilized in an
autoclave under the following conditions:
- Autoclave: Class B.
- Sterilization temperature: 134 °C.
- Time at sterilization temperature: 18 minutes.
19