Chapter 2
Design the Relative Standard Curve Experiment
Review the Reaction Setup
37
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
Complete the
Reaction Setup
Screen
To complete the Reaction Mix Calculations tab
for the c-myc assay
:
IMPORTANT!
Perform the following steps for each target assay in the reaction plate.
1.
In the Reaction Setup Screen, select the
Reaction Mix Calculations
tab (default).
2.
In the Select Target pane, select
c-myc
.
3.
In the Assay Type drop-down list, select
Inventoried/Made to Order
.
4.
Make sure
the Reaction Volume Per Well field displays
50 µL
.
5.
Make sure
the Excess Reaction Volume field displays
10%
.
6.
In the Reactions for c-myc pane:
a.
Make sure the Master Mix Concentration field displays
2.0
✕
.
b.
Make sure the Assay Mix Concentration field displays
20.0
✕
.
c.
Review the components and calculated volumes for the PCR reactions:
Component
Volume (µL) for 1 Reaction
Master Mix (2.0
✕
)
25.0
Assay Mix (20.0
✕
)
2.50
Sample (10
✕
) or Standard
5.0
‡
‡ The sample or standard volume is limited to 10% of the total reaction volume.
H2O
17.50
Total Volume
50.0
1
2
3
5
4
6b
6a
6c