Chapter 5
Analyze the Relative Standard Curve Experiment
View the Amplification Plot
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
92
Notes
If your experiment does not meet the guidelines above:
• Omit wells (see “Omit Wells from the Analysis” on page 105).
or
• Manually adjust the baseline and/or threshold (see “View the Analysis Settings” on
page 100).
For More
Information
For more information on the Amplification Plot screen, open the 7500 Software Help by
clicking or
pressing
F1
.
Baseline Examples
Baseline Set Correctly
The amplification curve begins after the maximum
baseline.
Baseline Set Too Low
The amplification curve begins too far to the right of the
maximum baseline. Increase the End Cycle value.
Baseline Set Too High
The amplification curve begins before the maximum
baseline. Decrease the End Cycle value.