![abi 7500 Getting Started Manual Download Page 154](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420154.webp)
Chapter 6
Design the Comparative C
T
Experiment
Review the Reaction Setup
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
130
Notes
Design
Guidelines
When you design your own comparative C
T
experiment:
• If you are using TaqMan reagents, select the type of assay you are using:
– Select
Inventoried/Made to Order
if you are using Applied Biosystems
TaqMan
®
Gene Expression Assays (Inventoried or Made to Order), or Applied
Biosystems Custom TaqMan
®
Gene Expression Assays.
– Select
Custom
if you are designing your own assays with Primer Express
®
software.
• Enter a number from 10 to 100 for the reaction volume/well. The 7500 system
supports reaction volumes from 20 to 100
µ
L. The 7500 Fast system supports
reaction volumes from 10 to 30
µ
L.
• Include excess reaction volume to account for the loss that occurs during pipetting.
Applied Biosystems recommends an excess reaction volume of at least 10%.
• Review the reaction mix concentrations for each target: If needed:
– For TaqMan reagents, edit the master mix and assay mix concentrations.
– For SYBR Green reagents, edit the master mix, forward primer, and reverse
primer concentrations.
– For 1-step RT-PCR, edit the reverse transcriptase concentration.
• Review the reaction mix components for each target:
– If you are running Fast PCR reactions, make sure you use Fast master mix in the
PCR reactions.
– If you are running standard PCR reactions, make sure you use standard master
mix in the PCR reactions.
– For 1-step RT-PCR, make sure you include reverse transcriptase in the PCR
reactions and use a specific buffer.
• Review the sample dilution calculations for each sample. If needed, edit the diluted
sample concentration (including units) and stock concentration.
For More
Information
For more information on:
• Completing the Reaction Setup screen – Open the 7500 Software Help by clicking
or pressing
F1
.
• Applied Biosystems assays – Refer to the:
–
TaqMan
®
Gene Expression Assays Protocol
–
Custom TaqMan
®
Gene Expression Assays Protocol
.