Chapter 6
Design the Comparative C
T
Experiment
Set Up the Run Method
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
126
Notes
Design
Guidelines
When you design your own comparative C
T
experiment:
• Enter a number from 10 to 100 for the reaction volume/well. The 7500 system
supports reaction volumes from 20 to 100
µ
L. The 7500 Fast system supports
reaction volumes from 10 to 30
µ
L.
• Review the thermal profile:
– Make sure the thermal profile is appropriate for your reagents.
– If you are performing 1-step RT-PCR, include a reverse transcription step.
If your experiment requires a different thermal profile, edit the thermal profile or
replace the run method with one from the Run Method library. The Run Method
library is included in the 7500 software.
For More
Information
For more information on:
• The Run Method library or on completing the Run Method screen – Open the
7500 Software Help by clicking
or pressing
F1
.
• Using Advanced Setup – See “Advanced Setup Workflow” on page 204.