![abi 7500 Getting Started Manual Download Page 112](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420112.webp)
Chapter 5
Analyze the Relative Standard Curve Experiment
View the Amplification Plot
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
88
Notes
5.
View the threshold values:
a.
In Plot Settings tab, select
Log
in the Graph Type drop-down list.
b.
Select
GAPDH
in the Target drop-down list.
c.
Select the
Threshold
check box to show the threshold (default).
d.
Verify that the threshold is set correctly.
In the example experiment, the
threshold is in the exponential phase.
5c
5a
5b