![abi 7500 Getting Started Manual Download Page 223](http://html.mh-extra.com/html/abi/7500/7500_getting-started-manual_451420223.webp)
Chapter 9
Analyze the Comparative C
T
Experiment
Analyze the Study
199
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
Notes
Analysis
Guidelines
When you review the Gene Expression Plot of a comparative C
T
study:
• Select the Technical Replicates tab or the Biological Replicates tab to organize and
view the sample data according to the associated technical replicate group or biological
replicate group.
• Change the endogenous control by clicking
Endo Control
, then selecting a new target.
• Change the reference sample by clicking
Ref Sample
, then selecting a new sample.
• Omit biological or technical replicates from the analysis. See “Omit Wells using the
Well Table” on page 200
• Display a subset of the data in the study by selecting one or more rows in the Technical
Replicates tab or the Biological Replicates tab, then by selecting
Hide unselected data
from plot
in bottom of the Gene Expression Plot.
Note:
Comparative Ct Example Study.edm does not contain biological replicate groups.
However, another example study that installed with the 7500 software better
demonstrates the use of biological groups. You can find the data file for the example
study on your computer at:
<drive>
:\Applied Biosystems\<
software name
>\experiments\
Comparative Ct Study Example (Biological Groups).edm
Omit Replicates from the Analysis
You can use the Technical Replicates and Biological Replicates tabs of the Gene
Expression screen to omit technical replicates and biological replicates from the analysis.
To omit a technical or biological replicate:
1.
In the navigation pane, select
Analysis
Gene Expression
.
2.
In the Gene Expression screen, select the
Technical Replicates
or
Biological
Replicates
tab according to the type of replicate that you want to omit.
3.
In the replicate table, scroll to the biological or technical replicate of interest, then
select the check box in the Omit column.
4.
Click
Analyze
when you are finished omitting wells.
IMPORTANT!
You cannot omit
all
technical replicates that belong to a reference sample,
belong to a reference biological group, or serve as the endogenous control for a study.
Note:
You can also omit the technical replicates in the Biological Replicate Details table
at the bottom of the Biological Replicates tab.
For More
Information
For more information on:
• Viewing the Gene Expression Plot – See “View the Gene Expression Plot and Well
Table” on page 162, or open the 7500 Software Help by clicking
or pressing
F1
.
• Calculating Relative Quantitation Values – Refer to the
User Bulletin #2: Relative
Quantitation of Gene Expression
.