Parameterization
80
FV4000 / FS4000
D184B097U02
Buttons
Menu item
Buttons Set parameter Buttons
Configurable
parameters
Comments
„Enter“ „Enter“
„Data“
„Step“
Display
1 large
1 small
Totalizer
Total_
Displays the total value of the totalizer block. The
displayed unit is UNIT_TOTAL.
„Step“
Totalizer
Status_
Displays the actual mode of the totalizer and the
status of the output variables (total status).
„Step“
Adr + State_
Displays the PA address and the status of cyclic
communication (Stop, Clear, Operate).
„Enter““
„Enter“
Display 2.row
Totalizer
Displays the flow totalizer.
„Enter““
„Enter“
Display 4.row
Q Operating
Mode
Q operating mode:
This operating mode is dependent upon the flow
display in operating volume or standard volume/mass
units.
„Enter“
Contrast
█████
The "Data" and "Step" buttons are used to set the
display contrast.
„Enter“ „Enter“
„Data“
„Step“
Error
Register
Mains
interrupt
14
Totalizer for the number of power failures since first
use.
„Data“
„Step“
. . . . . . . .
. . . . . . . .
Displays any errors that have occurred. Press the
"Enter" button to reset.
„Data“ „Enter“
„Data“
„Step“
Function
test
„Step“
Function
DIP-Switch
PA-Adr.:50
set by switch
„Enter“
1-7: Bus-Adr.
Protect
„Enter“
8: on = Adr by
swit
„Enter“
8: off = Adr by
bus
„Enter“
x = on - = off
xx------
Displays the current setting of the DIP switches on the
digital panel.
Switches 1-7:
PROFIBUS address
Switch 8:
Defines the addressing mode:
Off = Addressing via bus
On = Addressing via miniature switches 1-7
Note:
A change to the local bus address will only be applied
when the device is switched on.
Summary of Contents for FS4000
Page 1: ...Operating Instruction D184B097U02 Vortex Flowmeter Swirl Flowmeter FV4000 FS4000 ...
Page 116: ...Appendix 116 FV4000 FS4000 D184B097U02 ...
Page 117: ...Appendix D184B097U02 FV4000 FS4000 117 ...
Page 118: ...Appendix 118 FV4000 FS4000 D184B097U02 ...
Page 119: ...Appendix D184B097U02 FV4000 FS4000 119 ...
Page 120: ...Appendix 120 FV4000 FS4000 D184B097U02 ...
Page 125: ......