Setting items are as follows:
(1) Disable
Select to disable the function. The Modbus setting is 0.
(2) Weighing start
Select to assign the start weighing signal. The Modbus setting is 1.
(3) Weighing stop
Select to assign the stop weighing signal. The Modbus setting is 2.
(4) Weighing start/stop
Select to assign the start/stop weighing signal.
When the input signal is High (or Low), weighing will start and when the input signal is Low (or High),
weighing will stop. The Modbus setting is 3.
(5) Belt stop release
This is a function to undo the belt stop from external devices when the belt is stopped according to
“consecutive fail stop” or DO map settings.
Even if undone, the belt remains stopped. To start weighing again, press the START button.
The Modbus setting is 4.
(6) Summary clear
Select to assign clearing summary. The Modbus setting is 5.
(7) External NG1
Select to assign external NG1. The Modbus setting is 6.
(8) External NG2
Select to assign external NG2. The Modbus setting is 7.
(9) Metal detect
Select to assign metal detection. The Modbus setting is 8.
(10) DO1 to DO8 Forced behavior
Select to function DO forcibly. The Modbus settings of DO1 to DO8 are 9 to 16.