30
4.6.2.
Setting the target, high limit and lower limit values
The values used for detection are set.
1. Touch the Detect Basis button of the setting menu to open the screen below.
Fig.25 Detect Basis Setting screen
2. Select Target to enter the mass of the product to be weighed.
3. Select High Limit to enter the upper limit mass value of the product.
Here, enter the value deviated from the target value.
For example, when the target value is 100 g and the upper limit mass value is 150 g, enter 50 g in the High
Limit field. (150 g –100 g=50 g)
4. Select Lower Limit to enter the lower limit mass value of the product.
Here, enter the value deviated from the target value.
For example, when the target value is 100 g and the lower limit mass value is 50 g, enter 50 g in the Lower
Limit field. (100 g –50 g=50 g)
5. Touch the RETURN key
to return to the Product Setting screen.