5. Adjustment Operations
This chapter describes operations related to adjusting the AD-4961-2KD-2035 checkweigher.
5.1.
Calibrating the Checkweigher Using a Weight
The checkweigher calibration procedure is as follows:
1. Log in by the management level of Quality Manager or above.
2. In the Weighing screen, touch the Setting key
to open the Common Setting screen.
3. Select the Balance tab and touch the Cal button to open the Calibration screen.
Fig.50 Common
Setting screen (with Balance tab selected)
Fig.51 Calibration screen