73
8.4. DO
Behavior
8.4.1. Delay
time
Enter the time elapsed from judgment result output until a signal is output from the specified DO.
For details, refer to “4.6.6. Setting the delay and hold time.”
8.4.2. Hold
time
Enter the time a signal is being output from the specified DO.
For details, refer to “4.6.6. Setting the delay and hold time.”
8.4.3. Polarity
Set the polarity of a signal from DO.
Positive: When a signal is output from DO, the device on contact a turns on. The Modbus setting is 0.
Negative: When a signal is output from DO, the device on contact b turns on. The Modbus setting is 1.
8.5. Detect
Parameter
Displays the parameters used for weighing. The values will be changed automatically. Do not attempt to change them.
Only the A&D service personnel is allowed to do it.
Caution:
Changing the parameter values may cause the checkweigher malfunction.
8.6. Revision
8.6.1. Tare
value
The tare value is used to obtain the net weight of the product.
For details, refer to “4.6.4. Setting the tare value.”