76
8.9. Stability
Detect
8.9.1.
Stable mass range
Enter the mass range to detect stability.
8.9.2.
Stable time range
Enter the time range to detect stability.
8.10. Zero
Revision
8.10.1. Zero adjustment range
Enter the range to adjust zero.
When the actual mass value is not within the range, zero adjustment or zero tracking will not be performed.
8.10.2. Zero tracking mass range
Zero tracking is a function to perform zero adjustment automatically while the conveyor is stopped.
Enter the mass range to perform zero tracking.
The zero tracking mass range is 0.0 g, zero tracking will not be performed.
When the total mass value while the conveyor is stopped is within the zero range in a certain duration (zero tracking
time range), zero tracking will be performed.
8.10.3. Zero tracking time range
Enter the time range to perform zero tracking.
When the total mass value while the conveyor is stopped is within a certain range (zero tracking mass range) in the
zero tracking time range, zero tracking will be performed.
For example, when the zero adjustment range is 10 g, zero tracking mass range is 5 g and zero tracking time range
is 3 seconds, zero tracking will be performed when the actual mass value is 0±10 g and the total mass value is within
0±5 g for 3 seconds.
8.11. Calibration
For details, refer to “5.1. Calibrating the Checkweigher Using a Weight.”
8.11.1. Calibration weight value
Enter the calibration weight value to be used for calibration.
8.11.2. Zero
point
Enter the zero point value in mV/V.