Technical Specifications
Power Supply Output:
5.0 Vdc, 1200mA
Motor Voltage:
3.6 Volts DC Current without load:
1000 mA max.
Current with motor stalled:
3.5 amps max.
Charge Current:
500-1300 milliamps
Running Current:
700 mA
Charging Time:
4 hours initial, 4 hours thereafter
Battery Operating Time:
160 minimum minutes
Clipper Weight:
144 grams (5 oz.)
Clipper Dimensions:
16.1 cm (6.34 in) Height – with
9660
blade
17.5 cm (6.89 in) Height – with
9690
blade
4.1 cm (1.6 in) Width – with
9660
blade
5.1 cm (2.0 in) Width – with
9690
blade
4.1 cm (1.6 in) Width – 9661L body
14.2 cm (5.6 in) Height – 9661L body
12.7 cm (5.0 in) Circumference –
9661L
body
The device has been tested in accordance with ANSI/AAMI
ES60601-1:2005/(R) 2012, CAN/CSA-C22.2 NO. 60601-1:14, IEC
60601-1:2005 + A1: 2012 Medical Electrical Equipment – General
Requirements for Basic Safety and Essential Performance.
The device (including components and accessories) bearing the
CE mark complies with the requirements of the Medical Device
Directive 93/42/EEC, as amended per 2007/47EC.
Table 1 Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration tables - Electromagnetic Emissions
The
Model 9661L
is intended for use in the electromagnetic environment specified below. The customer or user of the
Model 9661L
device should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
Model 9661L
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment
RF emissions
CISPR 11
Class B
The
Model 9661L
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Table 2 Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The
Model 9661L
is intended for use in the electromagnet environment specified below. The customer or the end user of the
Model 9661L
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
+2 kV for power
supply lines
+1 kV for input/output lines
+2 kV for power
supply lines
+1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
+1 kV differential mode
+2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
>95% dip in 0.5 cycle
60% dip in 5 cycles 30%
dip for 25 cycles
>95% dip in 5 seconds
0% dip in 1 cycles
0% dip in 0.5 cycles@0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°
>95% dip in 0.5 cycle
60% dip in 5 cycles 30%
dip for 25 cycles
>95% dip in 5 seconds
0% dip in 1 cycles
0% dip in 0.5 cycles@0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the
Model 9661L
requires
continued operation during very long power
mains interruptions, it is recommended
that the
Model 9661L
be powered from an
uninterruptible power supply or battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.