SCANNING MODES AND SETTINGS
MAY 13, 2021
73
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Click the
button to display barcode label area. To rotate the label area by 180°, click
.
Figure 51: Barcode label area window
–
Undo / Redo modification made during specifying the
Scanmap
. These buttons
are inactive when
Use threshold
function is applied.
5.1.2.1
Focus Indicator
Click
to activate the
Focus indicator
function. Search for the sharpest focus on the
current field of view automatically by clicking
Auto focus
button.
To define focus limits, select
Limit the range
. Set the upper and lower limits by selecting the
desired value in the selection boxes. By using
± Range
, the
Lower limit
value is set to -32 of
the current focus value and the
Upper limit
value is set to +32 of the current focus value.
This is useful if
Auto Focus
does not find the focus on the specimen, but on the coverslip.
Also, this makes the focusing sequence faster.
For manual focusing define the manual focus step size by adjusting the
Step size
slider and
use the two arrow buttons to focus manually. This is useful when you are interested in an
area that remained blurry with
Auto Focus
(for example, the specimen is thicker in that
area). This does not influence the way slides are digitized in Auto focus mode.
Figure 52. Focus indicator panel