PANNORAMIC® SCAN RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
145
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth
with water.
•
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free
cotton cloth and gently rub the surface.
9.3.2
Cleaning the Magazines
NEVER
clean the magazines with
Xylene
because it might harm the plastic. We recommend
applying
Isopropanol
on a piece of lint-free cotton cloth and gently rub the surface instead.
In case the magazine is damaged, use either order a new one, as inserting a deteriorated
magazine could lead into unsafe loading of slides.
9.3.3
Preventive Maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage
temperature above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
9.4
Troubleshooting
This chapter describes the most common troubleshooting solutions for the Pannoramic®
SCAN RX product. If any problem arises that does not have solution in this manual, contact
3DHISTECH service support at
.
The product does not start.
•
Power supply is interrupted.
1.
Make sure that the 24V external power supply is on.
2.
Check cable connections.
Pannoramic® Scanner Software for Research control software has stopped.
•
Slide jammed
1.
Exit control software.
2.
Switch off control computer and 24V power supply, disconnect power cable.
3.
Remove protective covers. Follow the steps in chapter
4.
Remove slide manually (see chapter