PANNORAMIC® SCAN RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
147
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Slide is often positioned incorrectly.
•
Guiding fault in transport system
•
Check if faulty positioning disappears after the selection of a different type of
magazine (for example, universal magazine).
•
Contact 3DHISTECH Service for support.
Digitization process is extremely slow.
•
Extended focusing
option is activated.
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated.
•
Deactivate the
Z-stack
option if not required.
Image is out of focus. (in Manual mode)
•
Focus range limit is activated.
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect.
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide.
9.4.1
Fluorescent Applications (Optional)
The switching of the Lumencor SPECTRA 6 cannot be triggered, despite triggering connection
through the camera.
•
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
•
Lumencor SOLA SM is switched off.
•
Switch light source on, otherwise contact 3DHISTECH Service for support.