PANNORAMIC® SCAN RX 1.0
SETTINGS AND SERVICES
MAY 13, 2021
126
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
o
Set
Position
,
Objective type
, and
Magnification
settings of the objective.
o
Use manual camera changer
–
Changing the cameras is performed manually.
o
Pre-scan objective protection enabled
–
If activated, slide area near the edges will
not be scanned to record fast preview image. If inactive, Fast Preview processing
becomes slower.
o
Cameras
–
Activate
BF
and/or
FL
mode for the camera. The software automatically
loads the camera settings upon clicking the
Test and apply camera settings
button.
Preset profiles
are the following:
▪
Calibrated Linear
: Color scheme is defined by calibration (Gamma value is
not included).
▪
sRGB standard
: Color scheme is defined by calibration (standard Gamma
value -2.2- is applied).
▪
sRGB standard Microscope Like
: Color scheme is defined by calibration
(standard Gamma value / White Point: D50).
▪
Legacy
: A Color scheme calibrated for the previous software version is
applied.
▪
Custom
: parameters are freely customizable as follows:
▪
Color Scheme
▪
Calibrated
: Scheme is defined by the color calibration slide.
▪
Legacy
: A Color scheme calibrated for the previous
software version is applied.
▪
Raw
: the color information of the recorded raw image is
applied.
▪
White Point
▪
D65
: color temperature according to a standard illuminant
D65 white point value is applied (6500K / daylight).
▪
D50
: color temperature according to a standard illuminant
D50 white point value is applied (5000K / horizontal
daylight).
▪
Gamma
▪
Standard
: a value of 2.2 is used.
▪
Medium
: a value of 1.8 is used.
▪
Off
: a value of 1.0 is used.