PANNORAMIC® SCAN RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
67
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
•
Manual setting of Use marker with threshold
–
If this option is selected, only the
area encircled with a marker pen will be scanned. You can define the threshold level by
changing the threshold value in the
Marker threshold level
selection box in the range of
0-255. This sets the sensitivity of the specimen detection inside the marker area. To set the
desired threshold value manually, change the setting in the
Threshold
selection box in the
range of 0-255. The default value is 212.
•
Everything within marker
–
This function is available only to select if the
Use
marker with threshold
function is activated. To scan everything within the marked area,
check the
Scan everything within marker
selection box.
Figure 44. Scan everything within the marker selected / not selected
Dilate mask
–
Adjust the slider to enlarge or decrease the scanned area. The edge is
extended discretely by entering a new value each time. To switch to manual mode, click the
top left corner of the button and the required value can be entered in the box
.