Pannoramic MIDI II 2.0.5 User’s Guide
7 Troubleshooting and Maintenance
7.3.3 Installing and Removing the Filter Module (Fluorescent Option)
Figure 81 – Installing and removing the filter module
1. If using an Lumencor SOLA SM II illumination unit, turn it off.
Warning!
Crush hazard! Stop the running digitization process before you proceed
with filter module installation or removal.
2. Pull the pair of the protective cover halves apart or remove the right part completely (see
7.3.1 Opening the protective covers
), but do not switch the main unit off.
3. Select the desired filter and click
Set filter wheel to insert position
on the
Filter settings
panel (see section
5.6.1
) and the filter will be positioned in such a way that you will have free
access to the filter without having to turn the filter wheel manually.
4. Enter the respective parameters in the input fields.
Warning!
Crush hazard! Rotating the filter turret motor-operated during
installation or removal may lead to injury.
December 7, 2018 - Rev. 1