Pannoramic MIDI II 2.0.5 User’s Guide
7 Troubleshooting and Maintenance
2. The right half of the cover can be pulled apart only after the FL tray cover is removed. To
pull apart the other half, the left side FL tray cover must be removed first.
Note:
If necessary, one or both halves can be removed completely. For this
purpose, push them to the right or to the left until they can be
detached.
3. On completion of the maintenance or repair work, the cover halves must be pushed
together again and locked. A “clack” sound can be clearly heard.
Caution!
Make sure that no cables are squeezed as you replace the cover(s). This
refers in particular to the cable(s) of the scan camera.
7.3.2 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emerg. position
button.
Warning!
Crush hazard! Before you remove a jammed slide, cancel
the running
digitization process, terminate the software session and turn the power
off.
Any manipulation performed on a mechanical component must be followed by a complete device
initialization routine (software restart).