Pannoramic MIDI II 2.0.5 User’s Guide
7 Troubleshooting and Maintenance
Lumencor SOLA SM II illumination unit
•
Keep inflammable and combustible materials outside the range of
SOLA SM II
illumination
unit.
•
Do not expose yourself to the UV radiation of the
SOLA SM II
illumination unit.
•
When the
SOLA SM II
illumination unit is switched on:
◦
NEVER
look directly into the fiber-optic cable.
◦
NEVER
remove the fiber-optic cable from the device.
◦
NEVER
connect the fiber-optic cable to the device.
◦
NEVER
remove the housing from the device.
7.1.1 Cleaning the device
•
Before cleaning, switch off the device and pull the plug out.
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
•
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free
cotton cloth and gently rub the surface.
7.1.2 Cleaning the trays
NEVER
clean the trays with
Xylene
, because it might harm the plastic. We recommend to apply
Isopropanol
on a piece of lint-free cotton cloth and gently rub the surface instead. In case the tray is
damaged, use either order a new one, as inserting a deteriorated tray could lead into unsafe loading
of slides.
7.1.3 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
December 7, 2018 - Rev. 1
3DHISTECH Ltd.
124
(