Pannoramic MIDI II 2.0.5 User’s Guide
7 Troubleshooting and Maintenance
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Virtual slide is improperly focused almost across its entire surface
•
Focus range enabled
◦
Disable
Limiting focus range
function
Poor image quality
•
Imaging error (for example, chromatic aberrations)
◦
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
•
Scan camera is not properly adjusted (for example, due to loose fixture)
◦
Contact 3DHISTECH Service for support.
Slide is often positioned incorrectly
•
Guiding fault in transport system
◦
Contact 3DHISTECH Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
◦
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
◦
Deactivate the
Z-stack
option if not required.
December 7, 2018 - Rev. 1
3DHISTECH Ltd.
126
(
144
)